Cary 7000
Monochromator Double out-of-plane Littrow monochromator.
Grating
- 70 x 45 mm
- UV-Vis Ruled line diffraction grating, 1200 lines/mm blazed at 250 nm
- NIR Ruled line diffraction grating, 300 lines/mm blazed at 1192 nm
Beam splitting system
- Rotating beam splitter, which measures a sample, dark and reference signal
- per cycle with a speed of 30 Hz
Detectors
- UV-Vis R928 PMT
- NIR Cooled PbSmart PbS
Limiting resolution (nm)
- UV-Vis <0.048 nm
- NIR <0.2 nm
Stray light (%T)
- At 220 nm (10 g/L NaI ASTM method) <0.00007%
- At 370 nm (50 mg/L NaNO2) <0.00007%
- At 1420 nm (H2O, 1 cm pathlength) <0.0002%
- At 2365 nm (CHCl3, 1 cm pathlength) <0.00045%
Wavelength range (nm)
- (N2 purge required below 185 nm)
- 175–3300 nm
Wavelength accuracy (nm)
- Deuterium lamp lines
- UV-Vis: 190–900 nm +- 0.08 nm
- NIR: 760–3000 nm +- 0.4 nm
Wavelength reproducibility (nm)
- Peak separation of repetitive scanning of a UV-Vis line source <0.025 nm
- Peak separation of repetitive scanning of a IR line source <0.1 nm
- Standard deviation of 10 measurements, UV-Vis <0.005 nm
- Standard deviation of 10 measurements, NIR <0.02 nm
Photometric accuracy (Abs)
- Using double aperture method at 0.3 Abs UV-Vis <0.00025 Abs
Cary Universal Measurement Accessory
Measurement modes
- Absolute specular refl ection at variable angle from 5–85° in 0.02° intervals
- Direct transmission and variable angle transmission from 0–90° in 0.02° intervals
- Diffuse scattering, refl ection or transmission through independent sample rotation (360°) and detector
- positioning between 10–350° at 0.02° in intervals
- Absorptance, A where A= 1–R–T at variable angle without moving the sample or beam onto the sample
- for improved productivity and greater accuracy.
- Refl ection/Transmission at single wavelength (read) or wavelength range (scan)
Wavelength range 190–2800 nm
Auto polarizer wavelength range 250–2500 nm
Sample size
- Diameter:
- 5 mm minimum–275 mm maximum
- 255 mm if detector slide mount is installed
- 235 mm if a depolarizer is mounted in detector slide mount
- Maximum physical thickness: 30 mm using standard supplied sample holder
Apertures
- Incident beam: 1, 2 and 3 degrees
- Detector: 1, 1.8, 2, 3, 4, 4.4, 5 and 6°