Configuration: Reflected light microscope with trinocular eyepiece and chromatic aberration free infinity system.
Feature
Configuration
Objectives
Eyepiece 10x (F.N. 22)
NA
WD (mm)
Focal length (mm)
Physical depth of focus (mm)
Total magnification (M)
Actual field of view (Ømm)
Depth of focus (µm)
Plan Fluor BD 5x
0.15
18.0
40.0
12.22
50x
4.40
31.27
Plan Fluor BD 10x
0.30
15.0
20.0
3.06
100x
2.20
7.82
Plan Fluor BD 20x
0.45
4.5
10.0
1.36
200x
1.10
2.95
Plan Fluor BD 50x
0.80
1.0
4.0
0.43
500x
0.44
0.79
Plan Fluor BD 100x
0.90
2.0
0.34
1000x
0.22
0.50
Light source
50 W halogen lamp (Nikon LV-HL50W)
Filters
NCB11 (color balancing filter);
Polarizer & analyzer.
Camera
DS-Fi3 CMOS 5.9 Mpixel (6.91x4.92mm, 2880 x 2048 pixels), quant. 12 bit, sRGB 24bit. Exposure: 100us - 30 sec. Gain: 1.00 - 64.0x. Exposure compensation +1 EV by 1/6 EV step.
Software
NIS Elements D Camera control, measurements, annotations, HDR, large area acquisition using stitching, Z-series, macros, etc.
Sample stage
sample size up to 150x150x20 mm;
movement range 150x150x30 mm;
tracking & measuring capabilities (±1µm).