The Plasma System GIGAbatch 360M is a compact high-volume reactor for surface cleaning, hydrophilation, polymer & graphene isotropic etching, polymer activation (e.g., for bonding) and general oxidation on silicon, III/V-compound, quartz, ceramic, lithium niobate substrates.
Specifications:
Feature
|
Value/Configuration
|
Process chamber
|
Ø245 x 395 mm (18 Liters)
|
Process gas supply
|
MFC 1 gas: O2 (99.999%), 0-2000 sccm.
MFC 2 gas: empty ( ), 0-500sccm. Free slot for He, Ar, N2, CF4, etc.
|
Vent gas supply
|
VENT gas: N2 (99.999%), 0.5-1 bar.
|
Microwave generator
|
Power output: up to 1000 Watt
Frequency: 2.45 GHz
Coupling: Parabolic antenna
|
Vacuum system
|
Base pressure: < 0.1 mbar
Operating pressure: 0.6 – 1.5 mbar
Pump down time: typically, 1 min.
Valve: DN 40 KF Elbow type
|
Vacuum pump
|
Name: Edwards GX100N (located in ISSP’s basement)
Type: Dry pump
Pumping speed: up to 105 m3/h
Ultimate pressure: 5x10-3 mbar
|
Plasma system control
|
Plasma system gas valves are powered by compressed dry air (CDA) 4-6 bar. PC based system controller with 10.4” TFT color monitor, USB, Ethernet, RS232.
|
Process control
|
Chamber temperature monitor: IR sensor.
Plasma control: optical emission End Point Detection with plasma verification & vacuum set point as microwave interlock.
|
Sample holders
|
Horizontal quartz plate: 350x130x4 mm.
Quartz boat for 4” & 6” wafers, 27 slots.
|
Other accessories
|
Faraday cage.
|